Eonis MDRC-2224 WP Guia do Utilizador Página 30

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 29
7. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment
guidance
d=2.3P 800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
xed RF transmitters,
as determined by an
electromagnetic site
survey,
3
should be less
than the compliance
level in each frequency
range.
4
Interference may occur in
the vicinity of equipment
marked with symbol:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reection from structures, objects and people.
Recommended separation distance
The Eonis is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of the Eonis can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the Eonis as recommended below, according to the maximum output power of the communications
equipment.
3. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Eonis is used exceeds the applicable RF compliance level above, the Eonis should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Eonis.
4. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than3V/m.